Test Finger Force: Why 10 N and 30 N Are Not Interchangeable
Two force values dominate accessibility testing: 10 N and 30 N. Both are applied with the same family of probes, and the choice between them changes results on products with compliant covers, gaskets or thin walls.
Where the two values come from
10 N comes from the ingress protection standard, where the probe represents a finger exploring an opening. The intent is a deliberate but not forceful touch.
30 N comes from household appliance and plug/socket standards, where the probe represents a person pressing against a part to see whether it gives way. The intent is what an adult does when trying to reach something.
There is no 20 N level in either standard. The two values are not points on a scale; they are two different scenarios.
| Force | Source | Scenario |
|---|---|---|
| 10 N | IEC 60529 (IP) | exploring an opening |
| 30 N | IEC 60335-1, IEC 60884 | pressing to reach |
Why the difference changes results
A 10 N force on a compliant plastic cover deflects it slightly. A 30 N force deflects it much more — often enough to open a gap that the cover was designed to keep closed, or to bring the cover into contact with a live part behind it.
On a rigid cover the two forces give the same result. On a thin one they do not, and the thin ones are exactly the products where the question matters.
This is why a product can pass an IP2X check at 10 N and fail a 30 N accessibility check with the same probe. Neither test is wrong; they ask different questions.
Applying force repeatably
Hand pressure cannot produce 30 N consistently. A push-pull gauge or a calibrated spring is needed, and the force should be applied along the probe axis while the operator watches the reading, not applied by feel and then checked.
The duration matters as well. A momentary press does not let a compliant part reach its deflected position. The standard specifies a hold time, usually around a minute for the appliance case, and it exists because the deflection is time-dependent.
What the report should state
The force applied, the standard clause it came from, and whether the probe was held at that force for the specified duration. "Tested with test finger" without a force value is a result that cannot be compared with anything.
Where a product is subject to two standards with different forces, both results are worth recording. The 10 N pass and the 30 N fail together describe the product accurately; either alone misleads.
A common mistake
Applying 30 N with a probe that was designed for 10 N and has a small joint. The joint deflects under the higher force before the probe reaches its intended position, so the probe folds instead of pressing, and the result reads as "not accessible" when the geometry was never actually tested.
The probe must be capable of transmitting the force to its tip. On jointed probes this means applying force gradually and checking that the finger is not buckling.