Hipot Tester Daily Check: The Load Resistor, the 5 mA Lamp and the Ten Minutes That Save a Report
A hipot test produces a pass or a fail. A tester with a broken return lead produces a pass every time, because no current can flow. It is the most dangerous kind of instrument failure: silent, and in the direction of accepting bad product.
The check, in three steps
Connect a known resistor across the output, set a voltage, and confirm the current reading matches. A 1 MΩ resistor at 1 kV should read 1 mA; a 200 kΩ resistor should read 5 mA and should trip a 5 mA setting.
| Check | Setup | Pass criterion |
| Open circuit | No load | Reads zero current, no trip |
| Known load | 1 MΩ at 1 kV | 1 mA ± 5 % |
| Trip test | 200 kΩ at 1 kV, 5 mA trip | Trips within 1 s |
| Continuity of return lead | Short the leads | Reads full current, then trips |
| Display zero | No voltage | 0.0 kV, 0.00 mA |
Why 5 mA is the useful value
Most portable testers are set to trip at 5 mA for routine testing. Checking that the trip actually happens at that value, not just that the display shows current, is the part that catches a stuck trip circuit.
Record it
Write the date, the resistor value, the voltage, the current reading and the trip time on a log. Ten minutes a day produces a record that answers the question a customer will eventually ask: how do you know the tester was working on the day you tested my product.
Standards referenced: IEC 61010-1 for the safety of the tester itself, ISO/IEC 17025 clause 6.4 for equipment verification between calibrations, IEC 60243-1 for dielectric test methods.
Why a resistor and not a lamp
A lamp is non-linear: its resistance rises as it heats, so the current it draws depends on how long the test has been running. A resistor is linear, which makes the expected current predictable at any voltage.
Use a resistor rated for the voltage and power: at 1 kV and 1 MΩ, the dissipation is 1 W, so a 5 W resistor is adequate and a 1/4 W part will fail open after a few checks.
The check that catches a stuck trip
Set the trip to 5 mA and connect a 200 kΩ resistor. At 1 kV the current is 5 mA and the tester should trip within one second. A tester that shows 5 mA but does not trip has a faulty trip circuit, and every subsequent test result is suspect.
Frequency and record
Daily before the first test, and after any event that could have damaged the tester, such as a flashover or a dropped unit. Record the date, the resistor value, the voltage, the current reading and whether the trip occurred.
One line per day, twelve months a year: that is the evidence that the instrument was working when the product was tested.
Where the resistor is stored matters
A resistor kept in a drawer with metal tools collects debris and can flash over at 1 kV. Keep it in a small box with its value written on the lid, and inspect it monthly for cracks or discolouration.
Two resistors are better than one
Keep a 1 MΩ for the reading check and a 200 kΩ for the trip check. The first confirms the meter is linear; the second confirms the trip operates. A single resistor cannot do both well.
Record both readings. A tester that reads 1.02 mA on the 1 MΩ and trips in 0.4 s on the 200 kΩ is healthy; one that reads 1.00 mA but takes 3 s to trip has a slow comparator.