Julixing Instruments
 
Julixing Instruments

Company News

Hipot Tester Daily Check: The Load Resistor, the 5 mA Lamp and the Ten Minutes That Save a Report

Hipot Tester Daily Check: The Load Resistor, the 5 mA Lamp and the Ten Minutes That Save a Report

A hipot test produces a pass or a fail. A tester with a broken return lead produces a pass every time, because no current can flow. It is the most dangerous kind of instrument failure: silent, and in the direction of accepting bad product.

The check, in three steps

Connect a known resistor across the output, set a voltage, and confirm the current reading matches. A 1 MΩ resistor at 1 kV should read 1 mA; a 200 kΩ resistor should read 5 mA and should trip a 5 mA setting.

CheckSetupPass criterion
Open circuitNo loadReads zero current, no trip
Known load1 MΩ at 1 kV1 mA ± 5 %
Trip test200 kΩ at 1 kV, 5 mA tripTrips within 1 s
Continuity of return leadShort the leadsReads full current, then trips
Display zeroNo voltage0.0 kV, 0.00 mA

Why 5 mA is the useful value

Most portable testers are set to trip at 5 mA for routine testing. Checking that the trip actually happens at that value, not just that the display shows current, is the part that catches a stuck trip circuit.

Record it

Write the date, the resistor value, the voltage, the current reading and the trip time on a log. Ten minutes a day produces a record that answers the question a customer will eventually ask: how do you know the tester was working on the day you tested my product.

Standards referenced: IEC 61010-1 for the safety of the tester itself, ISO/IEC 17025 clause 6.4 for equipment verification between calibrations, IEC 60243-1 for dielectric test methods.

Why a resistor and not a lamp

A lamp is non-linear: its resistance rises as it heats, so the current it draws depends on how long the test has been running. A resistor is linear, which makes the expected current predictable at any voltage.

Use a resistor rated for the voltage and power: at 1 kV and 1 MΩ, the dissipation is 1 W, so a 5 W resistor is adequate and a 1/4 W part will fail open after a few checks.

The check that catches a stuck trip

Set the trip to 5 mA and connect a 200 kΩ resistor. At 1 kV the current is 5 mA and the tester should trip within one second. A tester that shows 5 mA but does not trip has a faulty trip circuit, and every subsequent test result is suspect.

Frequency and record

Daily before the first test, and after any event that could have damaged the tester, such as a flashover or a dropped unit. Record the date, the resistor value, the voltage, the current reading and whether the trip occurred.

One line per day, twelve months a year: that is the evidence that the instrument was working when the product was tested.

Where the resistor is stored matters

A resistor kept in a drawer with metal tools collects debris and can flash over at 1 kV. Keep it in a small box with its value written on the lid, and inspect it monthly for cracks or discolouration.

Two resistors are better than one

Keep a 1 MΩ for the reading check and a 200 kΩ for the trip check. The first confirms the meter is linear; the second confirms the trip operates. A single resistor cannot do both well.

Record both readings. A tester that reads 1.02 mA on the 1 MΩ and trips in 0.4 s on the 200 kΩ is healthy; one that reads 1.00 mA but takes 3 s to trip has a slow comparator.

    Standard:       IEC 61032 IEC 60061 IEC 60335 IEC 60529 IEC 60068 IEC 60695 VDE 0620 UL AS/NZS3112
    TAG:
IEC Test Probe Go No Go Gauge Spring Impact Hammer Integrating Sphere UL Probe Test IP Tester Test Equipment
 
Glow Wire Tester Impact Test Hammer Go Gauge UL Test Finger Probe Hipot Tester Test Probe Needle Impact Testing Machine
  Español   Español Svenska   Svenska لعربية   لعربية русский   русский Português   Português China  中国 Polski   Polski Română   Română Dansk   Dansk suomi   suomi
  Français   Français hrvatski   hrvatski Ελληνικά   Ελληνικά Nederlands   Nederlands česky   česky Deutsch   Deutsch Japan   日本の Italia   Italia български   български นไทย   นไทย
test probe