Leakage Current Measurement: The 2 kΩ Body Model and Why Your Meter Reads Something Else

Leakage Current Measurement: The 2 kΩ Body Model and Why Your Meter Reads Something Else

Leakage current is the current that would flow through a person touching an accessible part. That definition makes the instrument part of the measurement: the current depends on the impedance of the body, and the body is a network, not a resistor.

IEC 60990 defines the measuring networks. The most used is the one for perception and reaction: a 2 kΩ resistor in parallel with a capacitor, plus a series resistance, which together approximate the impedance of a hand touching metal at power frequency.

What the network does that a meter does not

QuantityThrough the 2 kΩ networkThrough a 10 MΩ multimeter
Reading at 230 V, 0.15 µF couplingAbout 0.22 mAVoltage only, no current path
Response to DC componentPartly blocked by the networkShown as a voltage
Response to high frequencyWeighted by the network capacitanceUnweighted
ResultTouch current as a person would feel itOpen-circuit voltage of a stray path

The last row is the common mistake. A multimeter with a high input impedance reads a voltage produced by a few nanoamps of capacitive coupling and reports 90 V, which is alarming and meaningless as a leakage figure.

The three networks that cover most standards

  • Perception and reaction. 2 kΩ with a parallel capacitor; used for accessible metal parts.
  • Let-go. Heavier weighting, used where a sustained current matters more than a brief shock.
  • Electric burn. Low impedance, used for high-current paths.

Quoting a current without naming the network is like quoting a weight without naming the planet. The report has to say which network, at which frequency, under which condition.

Practical measurement rules

Measure with the product in its worst normal condition, then repeat in each single-fault condition the standard requires. Ground the measuring network at the same point as the product, and keep the leads short: a 1 m loop picks up enough capacitive current at 50 Hz to matter when the limit is 0.25 mA.

Our leakage current testers implement the standard networks as plug-in modules, and the technical notes list which network belongs to which clause.

Standards referenced: IEC 60990 (measurement of touch current), IEC 60335-1 clause 13, IEC 61010-1 clause 6.3, IEC 62368-1 clause 5.4.

The 0.25 mA line and what it protects

For hand-held class II equipment the limit is often 0.25 mA at 50 Hz, which is roughly the threshold at which a person notices a tingle. A reading of 0.4 mA is not dangerous, but it is over the line, and the fix is usually 2 nF of stray capacitance between a switching supply and an accessible metal part.

Measure with the network, not with a meter, and measure at the worst supply voltage within the rating.

For context, 0.25 mA at 50 Hz is below the let-go threshold by more than an order of magnitude, and a 2 kΩ body model at 230 V would pass 115 mA if the path were direct, which is why the network, the capacitor and the 0.15 µF coupling value all matter to the number in the report.

    Standard:       IEC 61032 IEC 60061 IEC 60335 IEC 60529 IEC 60068 IEC 60695 VDE 0620 UL AS/NZS3112
    TAG:
IEC Test Probe Go No Go Gauge Spring Impact Hammer Integrating Sphere UL Probe Test IP Tester Test Equipment
 
Glow Wire Tester Impact Test Hammer Go Gauge UL Test Finger Probe Hipot Tester Test Probe Needle Impact Testing Machine
  Español   Español Svenska   Svenska لعربية   لعربية русский   русский Português   Português China  中国 Polski   Polski Română   Română Dansk   Dansk suomi   suomi
  Français   Français hrvatski   hrvatski Ελληνικά   Ελληνικά Nederlands   Nederlands česky   česky Deutsch   Deutsch Japan   日本の Italia   Italia български   български นไทย   นไทย
test probe