Hipot Tester Daily Verification: One Kilovolt, Five Milliamps and the Load Resistor

Hipot Tester Daily Verification: One Kilovolt, Five Milliamps and the Load Resistor

A hipot tester has one job: apply a voltage and detect when the current through the insulation exceeds a limit. IEC 61180 sets out how the voltage is generated, measured and verified for low-voltage equipment, and that is the framework behind every check described here. Both halves of the tester's job can fail independently, and both fail in the direction that passes product. The voltage can be low while the display reads correct, and the trip point can be high while the instrument still buzzes and flashes at the end of a test.

The three quantities to check

Output voltageSet 1 kV, measure at the probes with a divider or a high-voltage meter. A 5 percent error is common after a repair and is invisible on the tester's own display.
Trip currentSet 5 mA, apply a resistor that draws slightly more, confirm the tester trips. Apply one that draws slightly less, confirm it does not.
Ramp and dwellTime the ramp to the set voltage and the dwell at voltage. A fast ramp stresses insulation differently from the specified one.

The resistor for the trip check is chosen by Ohm's law: at 1 kV, 5 mA corresponds to 200 kΩ. A 180 kΩ resistor draws about 5.6 mA and should trip a 5 mA setting; a 220 kΩ resistor draws about 4.5 mA and should not.

Why a resistor, not a lamp

A lamp or a shorting link tells you only that the tester trips at some current above zero. A resistor of known value tells you where it trips. The difference matters because most real failures are marginal: a product with 4.8 mA of leakage on a 5 mA limit is either good or bad depending on whether the tester actually trips at 5.0 mA or at 6.5 mA.

Two practical points about the resistor:

  • Use a high-voltage resistor rated well above the test voltage. A 1 kV test with a 500 V resistor will flash over inside the body and read as a trip, which looks like a pass.
  • Mount it in a probe housing with insulated leads. Holding a resistor in the hand while the tester ramps is how the check becomes a hazard.

The failures that daily verification catches

  1. Voltage divider drift. The feedback divider that sets the output drifts with age and temperature. The display still reads 1 kV and the output is 850 V. Product that should have been stressed at 1 kV passes at 850 V, with no indication in the record.
  2. Trip threshold drift. The current sense circuit ages, and the trip point moves up. A 5 mA limit becomes an effective 7 mA. Marginal product passes.
  3. Ramp too fast. A repaired instrument with a different ramp setting charges capacitive loads quickly, producing a current spike that trips on good product. The operator notices and raises the limit, and the limit stays raised.

The third is the most corrosive because it changes the test permanently to make the tester usable.

A five-minute daily routine

  1. Short the probes together, start a test at the working voltage and confirm an immediate fail. This checks that the detection path works at all.
  2. Connect the trip resistor and confirm a fail at the set current. Confirm the displayed trip current is within tolerance of the resistor's calculated value.
  3. Connect a resistor of about 20 percent higher value and confirm a pass, with the leakage reading close to the calculated value.
  4. Measure the output voltage with an external meter at the working voltage.
  5. Record the three readings, not a tick. Three numbers per day, thirty seconds each, produce a trend that shows drift before it becomes a failure.

What an annual calibration does not cover

An annual calibration certificate confirms that the instrument was within tolerance on the day it was calibrated. It says nothing about the intervening 364 days. Between calibrations, the only evidence is the daily check record, and a tick box with no number is not evidence.

Two further gaps are worth closing:

  • Probe and lead condition. A cracked high-voltage lead leaks current to the bench, which reads as product leakage. Inspect the leads weekly and keep a spare set.
  • Return-to-zero. After a test the output should discharge within a defined time. A tester that holds charge on the probes is a shock hazard and a source of repeat readings that look like leakage.

Choosing a limit that means something

The trip current is not a property of the tester; it comes from the product standard and depends on the application. What matters for the tester is that the setting can be verified and that the resolution is fine enough to be useful. A tester with a 5 mA setting and a 0.1 mA resolution gives a reading the operator can trend; a tester that only reports pass or fail gives nothing.

The electrical safety test range includes hipot testers with verifiable trip points and calibrated high-voltage dividers; the site index links the probe sets and discharge accessories used with them.

    Standard:       IEC 61032 IEC 60061 IEC 60335 IEC 60529 IEC 60068 IEC 60695 VDE 0620 UL AS/NZS3112
    TAG:
IEC Test Probe Go No Go Gauge Spring Impact Hammer Integrating Sphere UL Probe Test IP Tester Test Equipment
 
Glow Wire Tester Impact Test Hammer Go Gauge UL Test Finger Probe Hipot Tester Test Probe Needle Impact Testing Machine
  Español   Español Svenska   Svenska لعربية   لعربية русский   русский Português   Português China  中国 Polski   Polski Română   Română Dansk   Dansk suomi   suomi
  Français   Français hrvatski   hrvatski Ελληνικά   Ελληνικά Nederlands   Nederlands česky   česky Deutsch   Deutsch Japan   日本の Italia   Italia български   български นไทย   นไทย
test probe